What does 'QUALIFICATION CONTRACT REQUESTED' mean on Millipore Elix 35 Clinical Laboratory Equipment?
- MMaureen DavidSep 7, 2025
The message 'QUALIFICATION CONTRACT REQUESTED' indicates that a requalification of the validation is needed.
What does 'QUALIFICATION CONTRACT REQUESTED' mean on Millipore Elix 35 Clinical Laboratory Equipment?
The message 'QUALIFICATION CONTRACT REQUESTED' indicates that a requalification of the validation is needed.
What does 'SERVICE CONTRACT REQUESTED' mean on Millipore Laboratory Equipment?
The message 'SERVICE CONTRACT REQUESTED' means that a timer has indicated that a Service Contract visit is due.
How to resolve 'DIST WATER CONDUCTIVITY < MINI' error on Millipore Laboratory Equipment?
Keep the system operating for a few minutes to remove any air that could be trapped in the Resistivity Sensor.
What does 'PRODUCT CONDUCTIVITY < MINI' mean on Millipore Elix 35 Clinical, and how do I resolve it?
To address the 'PRODUCT CONDUCTIVITY < MINI' message, keep the System in PRODUCTION Mode for a few minutes to remove any air that could be trapped in the Resistivity Sensor.
What does 'PERMEATE < MINI' mean on Millipore Laboratory Equipment, and how can I resolve it?
If you see 'PERMEATE < MINI', verify that the feedwater is not already purified. Very low feedwater conductivity can lead to very low permeate conductivity values.
Brand | Millipore |
---|---|
Model | Elix 35 Clinical |
Category | Laboratory Equipment |
Language | English |
Guide for normal operation and maintenance of Elix 35/70 Clinical systems.
Important safety instructions and symbols for operating the Elix Clinical System.
Contact information for Millipore SAS via internet and manufacturing site.
Overview of Elix Clinical Systems designed for NCCLS Type II and Type I water.
Details feed water requirements and product water quality for Elix systems.
Lists dimensions, weight, environmental, electrical, and noise level specifications.
Diagram illustrating the main components of the Elix Clinical System.
Detailed explanations of each main component within the Elix Clinical System.
Instructions for filling the reservoir using automatic and manual methods.
Explanation of different operating modes like STANDBY, PRODUCTION, etc.
Guide to viewing system performance parameters via the DATA button.
How to find system serial numbers, catalogue numbers, and consumable information on the LCD.
Overview of scheduled maintenance tasks for system components and their frequency.
Step-by-step guide for removing and installing the Progard TL pack.
Procedure for replacing the Q-Gard TL Polisher Pack.
Instructions for replacing the Opticap final filter.
Information on replacing the UV Lamp, recommended to use a service representative.
Methods for sanitizing RO cartridges, both automatic and manual.
Procedures for cleaning RO cartridges using acid or base cleaners.
Classification and display of system messages including maintenance, alarm, and alarm stop.
Details on specific LCD messages, their meaning, and required actions.
List of consumable items with catalogue numbers and comments.
List of accessories with catalogue numbers and comments.
Catalogue numbers for different Elix Clinical System configurations.
Checklist to verify all items are present before system installation.
Checklist to confirm proper installation and configuration of the Elix Clinical System.
Procedure for using the emergency by-pass port for feedwater.
Log sheet for recording daily maintenance and operating parameters.