Zero Calibration
Run the conveyor for several minutes to warm up the belt and insure that
it is empty. Test weights are not used during a zero calibraion.
Perform Span Calibration,
see next page
initial zero count
zero calculation,
frequency count displayed
zero deviation calculated,
initial deviation = 0
zero calibration required
zero deviation accepted,
initial zero count = ###
Record this value for
future reference
START UP
PL-531 31