P940 load cell ’A’, mV
in
Displays the mV signal input from the load cell. Range 0.00 - 60.00 mV.
P941 load cell ’B’, mV
in
Displays the mV signal input from the load cell. Range 0.00 - 60.00 mV.
P942 V/F converter, V
in
Displays the input voltage to the voltage to frequency converter.
Range 0 - 3.98 V
P943 V/F converter, f
out
Displays the output frequency of the voltage to frequency converter.
Range 0 - 131,072
P944 power sensor
Displays a voltage supply reference for diagnostic purposes.
P949 diagnostic errors
Enables or disables diagnostic error checking, E101 - E104
entry: 0 = disable
1 = enable
Refer to Troubleshooting.
P950 zero register
Registers the number of zero calibrations that have been done since
the last master reset.
97/07/07
PARAMETER DESC.
PL-531 74