To replace the batteries unscrew the battery
compartment cap and replace all four 1.5V batteries
while paying attention to their polarity. Make sure
the gasket is in place before screwing back the cap.
Batteries should only be replaced in a
nonhazardous area using the battery type specified
in this instruction manual.
ACCESSORIES
MI56P Replaceable electrode for pH55 & pH56
M10004B pH 4.01 buer, 20 ml sachet, 25 pcs.
M10007B pH 7.01 buer, 20 ml sachet, 25 pcs.
M10010B pH 10.01 buer, 20 ml sachet, 25 pcs.
MA9004 pH 4.01 buer, 230 ml bottle
MA9006 pH 6.86 buer, 230 ml bottle
MA9007 pH 7.01 buer, 230 ml bottle
MA9009 pH 9.18 buer, 230 ml bottle
MA9010 pH 10.01 buer, 230 ml bottle
MA9015 Electrode storage solution, 230 ml
MA9016 Electrode cleaning solution, 230 ml
M10000B
Electrode rinse solution, 20 ml sachet,
25 pcs.
SPECIFICATIONS
RANGE -2.0 to 16.0 pH (pH55);
-2.00 to 16.00 pH (pH56);
-5.0 to 60.0°C / 23.0 to 140.0°F
RESOLUTION 0.1 pH (pH55);
0.01 pH (pH56);
0.1°C / 0.1°F
ACCURACY (@25°C) ±0.1 pH (pH55);
±0.05 pH (pH56);
±0.5°C / ±1°F
TYPICAL EMC DEVIATION ±0.1 pH (pH55);
±0.02 pH (pH56);
±0.3°C / ±0.6°F
TEMPERATURE COMPENSATION Automatic,
from -5 to 60°C
CALIBRATION Automatic, 1 or 2 points
with 2 sets of memorized buers
(pH 4.01, 7.01, 10.01 or 4.01, 6.86, 9.18)
PROBE Replaceable MI56P
ENVIRONMENT -5 to 50°C; 100% RH max.
BATTERY TYPE 4 x 1.5V; IEC LR44, A76
BATTERY LIFE approx. 300 hours of use
AUTO-OFF after 8 minutes of non-use
DIMENSIONS 200 x dia 38 mm
WEIGHT 100 g
CERTIFICATION
Milwaukee Instruments conform to the CE
European Directives.
Disposal of Electrical & Electronic Equipment.
Do not treat this product as household waste. Hand
it over to the appropriate collection point for the
recycling of electrical and electronic equipment.
Disposal of waste batteries. This product contains
batteries. Do not dispose of them with other
household waste. Hand them over to the appropriate
collection point for recycling.
Please note: proper product and battery disposal
prevents potential negative consequences
for human health and the environment. For
detailed information, contact your local
household waste disposal service or go to
www.milwaukeeinstruments.com (USA & CAN)
or www.milwaukeeinst.com.
RECOMMENDATION
Before using this product, make sure it is entirely
suitable for your specific application and for the
environment in which it is used. Any modification
introduced by the user to the supplied equipment
may compromise the testers performance. For your
and the tester’s safety do not use or store the tester in
hazardous environment. To avoid damage or burn, do
not perform any measurement in microwave ovens.
USER MANUAL
pH55 & pH56
PRO Waterproof
pH/Temperature
Testers
pH56
ON/OFF
CAL
SET
HOLD
pH56 PRO
pH/Temp
Waterproof
ISTPH56 11/20
SATISFACTION
GUARANTEED

Other manuals for Milwaukee PH55

Question and Answer IconNeed help?

Do you have a question about the Milwaukee PH55 and is the answer not in the manual?

Summary

Product Accessories

Certification and Disposal Information

Technical Specifications

Usage Recommendations

Operational Guide

Calibration Procedure

Setup Options

Electrode Replacement

Battery Replacement

Summary

Product Accessories

Certification and Disposal Information

Technical Specifications

Usage Recommendations

Operational Guide

Calibration Procedure

Setup Options

Electrode Replacement

Battery Replacement

Overview

The Milwaukee pH55 and pH56 PRO Waterproof pH/Temperature Testers are handheld devices designed for accurate measurement of pH and temperature. These testers are built to be robust and are suitable for various applications, including those requiring high accuracy and operation in challenging environments.

Function Description:

The primary function of these testers is to measure pH and temperature simultaneously. The pH value is displayed on the main LCD, while the temperature is shown on the secondary LCD. The devices feature automatic temperature compensation (ATC) to ensure accurate pH readings across varying temperatures. They also incorporate an instability indicator (hourglass) that turns off once the reading stabilizes, providing confidence in the measurement. The testers are equipped with a Battery Error Prevention System (BEPS) to prevent erroneous readings due to low battery levels by automatically switching off the device when batteries are weak.

Important Technical Specifications:

  • pH Range:
    • pH55: -2.0 to 16.0 pH
    • pH56: -2.00 to 16.00 pH
  • Temperature Range: -5.0 to 60.0°C / 23.0 to 140.0°F
  • pH Resolution:
    • pH55: 0.1 pH
    • pH56: 0.01 pH
  • Temperature Resolution: 0.1°C / 0.1°F
  • pH Accuracy (@25°C):
    • pH55: ±0.1 pH
    • pH56: ±0.05 pH
  • Temperature Accuracy (@25°C): ±0.5°C / ±1°F
  • Typical EMC Deviation:
    • pH55: ±0.1 pH
    • pH56: ±0.02 pH
    • Temperature: ±0.3°C / ±0.6°F
  • Temperature Compensation: Automatic, from -5 to 60°C
  • Calibration: Automatic, 1 or 2 points with 2 sets of memorized buffers (pH 4.01, 7.01, 10.01 or 4.01, 6.86, 9.18)
  • Probe: Replaceable MI56P
  • Environment: -5 to 50°C; 100% RH max.
  • Battery Type: 4 x 1.5V; IEC LR44, A76
  • Battery Life: Approx. 300 hours of use
  • Auto-Off: After 8 minutes of non-use
  • Dimensions: 200 x dia 38 mm
  • Weight: 100 g

Usage Features:

  • Activation and Measurement: To begin, the electrode needs to be activated by immersing it in MA9015 storage solution for 2 hours. The tester is turned on by pressing the ON/OFF button. After immersing the electrode in the solution to be tested, gentle stirring and waiting for the instability indicator to turn off will provide a stable reading.
  • Display: The main LCD shows the pH value, while the secondary LCD displays the temperature. A "CAL" tag indicates when the tester is calibrated.
  • Hold Function: In measurement mode, pressing and holding the SET/HOLD button freezes the display, showing "HOLD" on the secondary display. Any button press returns to normal mode.
  • Auto-Off: The device automatically turns off after 8 minutes of non-use to conserve battery life. An "OFF" message appears on the secondary display before shutdown.
  • Calibration Procedure:
    • Entering Calibration: Press and hold the ON/OFF/CAL button until "CAL" appears on the secondary LCD.
    • Buffer Recognition: The tester automatically recognizes the buffer solution (e.g., pH 7.01, 4.01, 10.01). If an invalid buffer is detected, "WRNG" is displayed.
    • Single-Point Calibration: Immerse the electrode in any valid buffer. If using pH 7.01 (or 6.86), press SET to exit, and "OK 1" confirms the calibration.
    • Two-Point Calibration: Recommended for better accuracy. Start with pH 7.01 (or 6.86), then rinse and immerse in the second buffer (pH 4.01, 10.01, or 9.18). "OK 2" confirms successful two-point calibration.
    • Exiting Calibration: Press ON/OFF during calibration mode to return to normal measurement without saving changes ("ESC" is displayed).
    • Reset Calibration: Press SET/HOLD after entering calibration mode (before the first point is accepted) to reset to default values and clear previous calibration ("CLR" is displayed).
  • Setup Mode: Allows selection of temperature unit (°C or °F) and pH buffer set (pH 7.01 BUFF or pH 6.86 BUFF for NIST). Enter setup by pressing ON/OFF until "TEMP" is displayed. Use SET/HOLD to change settings and ON/OFF to navigate or exit.

Maintenance Features:

  • Electrode Care: After use, the electrode should be rinsed with water to minimize contamination and stored with a few drops of MA9015 storage solution or pH7 (M10007) solution in the protective cap. Distilled or deionized water should never be used for storage.
  • Electrode Replacement: The MI56P electrode is replaceable. To replace, unscrew the plastic ring on top, pull out the old electrode, and insert a new one, ensuring gaskets are in place before screwing the ring back.
  • Battery Replacement: When the battery symbol lights up on the LCD, it indicates weak batteries and advises replacement. The tester uses four 1.5V IEC LR44 (A76) batteries. To replace, unscrew the battery compartment cap, replace the four batteries, paying attention to polarity, and ensure the gasket is in place before re-screwing the cap. Batteries should only be replaced in a non-hazardous area.
  • Calibration Frequency: Frequent calibration is recommended for better accuracy, especially after electrode replacement, testing aggressive chemicals, or when extreme accuracy is required.
  • Accessories: A range of buffer solutions (pH 4.01, 7.01, 10.01, 6.86, 9.18), electrode storage solution (MA9015), electrode cleaning solution (MA9016), and electrode rinse solution (M10000B) are available to support the maintenance and calibration of the device.
  • Disposal: The product and its batteries should not be treated as household waste. They should be handed over to appropriate collection points for recycling electrical and electronic equipment and waste batteries, respectively, in accordance with CE European Directives.

Milwaukee PH55 Specifications

General IconGeneral
BrandMilwaukee
ModelPH55
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals