EasyManua.ls Logo

MIMO UM-710S - Page 16

MIMO UM-710S
21 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PAGE 16 OF 21
Calibration
The product is with factory default calibration value. If required user may execute Touch Screen Controller
Calibration as follow.
Calibration is very important procedure to align coordinates of the exact touch point on screen with Windows
cursor. With the calibration user may utilize Touch Screen Control without distortion.
The last calibration data will be saved in the Windows system so as not to be required after reboot the system
except the user rotate the screen.
1.1 Click Start->Programs->UPDD->Setting
1.2. Click UPDD Consol (Pointer Device Setting) icon on system tray ->Calibration
Touch + mark position in order thenConfirm
If calibration point inputs are correct click Confirm
In case any mistake in calibration points inputs click Cancelor leave the screen untouched for about 10
seconds then the Calibration Data will be scrapped.
Calibrated Touch Screen coordinates may vary upon the lapse of usage time. Periodical
calibration is recommended.