Product Specifications
2-3
2.4 Environment Requirements
Operating environment, storage environment and running environment
Table 2-3 Overall environment requirements
Operating Environment
Requirements
Storage Environment
Requirements
Running Environment
Requirements
Ambient
Temperature
10℃~30℃ -10℃~40℃
10℃~40℃
Relative
Humidity
20%~85% 10%~90%
10%~90%
Atmospheric
Pressure
70kPa~106kPa 50kPa~106kPa
70kPa~106kPa
2.5 Product Specifications
2.5.1 Sample mode
Two sample modes are provided: whole blood mode, and prediluted mode.
2.5.2 Throughput
The throughput of BC-20s in OV-WB/OV-PD is no less than 40 samples/hour;
The throughput of BC-30s in OV-WB/OV-PD is no less than 70 samples/h;
2.6 Testing Parameters
The analyzer provides quantified results for 21 report parameters (WBC, RBC, PLT,
HGB, etc.) and 3 histograms (WBC, RBC, and PLT). See the table below for details.
Table 1 Parameters
Name Abbreviation
White Blood Cell count WBC
Lymphocyte number Lymph#
Mid-sized Cell number Mid#
Granulocyte number Gran#
Lymphocyte percentage Lymph%
Mid-sized Cell percentage Mid%
Granulocyte percentage Gran%
Red Blood Cell count RBC
Hemoglobin concentration HGB
Mean Corpuscular Volume MCV
Mean Corpuscular Hemoglobin MCH