What does Rotation Motor Error mean on Mindray Measuring Instruments?
- XxloganSep 12, 2025
If the Mindray Measuring Instruments displays a 'Rotation Motor Error', it could be due to a jammed sample probe, poor contact of the signal line, a damaged motor, poor connection between the drive board and the CUP board, or a malfunctioning photo coupler. Open the front panel and check if the sample probe is jammed. Enter the “Service ? System Test” screen and check the motor. The error will be removed if the test result is normal.