Customizing the Analyzer Software
5-31
5.11 Parameter Units
The “Parameter Units” screen is where you view and/or set (if you have the administrator
password) the reporting units of the parameters.
See Table 5-2 for the available units for every parameter groups. Note that if you choose g/L
or g/dL for the HGB/MCHC group, the MCH unit will automatically change to pg and its
reporting format will be ***.*; if you choose mmol/L for the HGB/MCHC group, the MCH unit
will automatically change to fmol and its reporting format will be **.**.
Table 5-2 Reporting unit
Parameter group Reporting format Reporting unit Remarks
***.*
10
9
/L Default
***.*
10
3
/uL
****
10
2
/uL
WBC
Lymph#
Mid#
Gran#
***.*
/nL
Lymph%
Mid%
Gran%
**.*
%
Default
***
g/L Default
**.*
g/dL
HGB
MCHC
**.*
mmol/L
*.**
10
12
/L Default
*.**
10
6
/uL
***
10
4
/uL
RBC
*.**
/pg
**.*
% Default
HCT
.***
L/L
***.*
fL Default
MCV
RDW-SD
***.*
um
3
RDW-CV **.*
% Default
****
10
9
/L Default
****
10
3
/uL
***.*
10
4
/uL
PLT
****
/nL
MPV **.*
fL Default

Table of Contents