Troubleshooting Your Analyzer
11-2
11.2 Errors without available error messages
Error Possible Cause(s) Recommended Action
The analyzer cannot
be turned on.
1. The power cord is
broken or not well
connected;
2. The fuse is broken;
3. The power outlet has no
electricity.
1. Check the power cord connection;
2. Check the fuse;
3. Check the electrical outlet.
Liquid drips from
analyzer inside.
Damaged pump hose or
blocked filter.
1. Turn off the power and wipe the
analyzer dry.
2. Call Mindray customer service
department or your local distributor
for assistance.
Recorder does not
work.
1. Recorder paper is
jammed;
2. Something is wrong with
the circuit.
1. Remove the jammed paper.
2. If the problem remains, turn off the
analyzer and turn it on again in 10
seconds.