Maintaining Your Analyzer
10-33
10.9 Calibrating Sample Probe Position
z Samples, controls, calibrators and waste are potentially infectious. Wear
proper personal protective equipment (e.g. gloves, lab coat, etc.) and follow
safe laboratory procedures when handling them in the laboratory.
z The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
The relative position between the sample probe and probe wipe block has influence on the
analysis results. In the accessory box, there is a sample probe localizer, as Figure10-41
shows. You need to use the localizer to adjust the position of the sample probe if you have
replaced wipe block, or observed motor error, or wrong analysis result. Also, as required by
regular maintenance, you should use the localizer to adjust the position of the sample probe
monthly.
Figure10-41 Sample probe localizer
Follow the steps below to do so:
1. SELECT Setup Password” and enter the administrator password (3000);
2. Push the right door latch in the direction indicated in Figure10-42;