Customizing the Analyzer Software
5-6
5.3 Count time
The “Count Time” screen is where you view and/or set (if you have the administrator
password) the reference time for the WBC and/or RBC count portion of the measurement
cycle. If the actual WBC or RBC count time (see Chapter 3.4.1 and 3.5.1) deviates from the
reference time by 2 seconds or more, the analyzer will alarm you for clogging or bubbles and
invalidate the results of all related parameters.
5.3.1 Entering the “Count Time” screen and viewing the
settings
Press [MENU] to enter the system menu.
Figure 5-7 System menu
SELECTSetup Count TimeFigure 5-7to enter the Count Time screen (Figure 5-8).