EasyManua.ls Logo

Mindray BC-3000 Plus - 8.5 “X-B Analysis” Program

Mindray BC-3000 Plus
325 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Using the QC Programs
8-39
8.5 “X-B Analysis” Program
The X-B
analysis is a weighted moving average analysis that uses values obtained from
patient samples. It was proposed by Brian Bull, M.D. using the 3 red cell indices, MCV, MCH
and MCHC to indicate the hematology instrument performance. Effective use of X-B requires
randomization of samples and a normal cross section of patients to prevent skewing of
indices.
It is recommended the X-B analysis be enabled when the sample volume of your laboratory is
greater then 100 samples per day.
8.5.1 Editing X-B Settings
Entering the “X-B Edit” screen
Press [MENU] to enter the system menu.
Figure 8-46 System menu
SELECT Quality Control X-B Analysis Limit” (Figure 8-46) to enter the “Limit
screen (Figure8-47).