12-6 Probes and Biopsy
Procedures for operating (with no biopsy function)
Inspection before examination
Connection to the ultrasonic
diagnostic system
Examinations
Wiping off the ultrasound gel
Washing the
transducer with water
Draining/drying
Immersion into disinfectant
Removing the transducer
from disinfectant
Rinsing the transducer into sterile water
Draining/drying
Inspection after use
Storage
Disinfection
Disconnection to the ultrasonic
diagnostic system
WARNING:
Disinfect the probe and sterilize the needle-guided bracket
before and after an ultrasound-guided biopsy procedure is
performed. Failure to do so may cause the probe and the
needle-guided bracket become source of infection.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Z50 and is the answer not in the manual?

Mindray Z50 Specifications

General IconGeneral
Battery BackupYes
Power Supply100-240V AC, 50/60Hz
Display Size15 inches
Display TypeLCD
Imaging ModesColor Doppler
ConnectivityDICOM, USB, Ethernet
ApplicationsObstetrics, Gynecology, Cardiac, Vascular, Urology, Musculoskeletal
Storage500GB HDD
Operating SystemWindows-based