Minebea Intec S-Type Load Cell LC Tigo PR 76 - Ordering Information; Technical Data

Default Icon
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.5 Ordering information
Model Max. capacity
E
max
Type
PR 76/60 kg 60 kg ..N/C3
PR 76/125 kg 125 kg ..N/C3
PR 76/250 kg 250 kg ..N/C3
PR 76/500 kg 500 kg ..N/C3
PR 76/1 t 1 t ..N/C3
PR 76/2 t 2 t ..N/C3
PR 76/3 t 3 t ..N/C3
PR 76/5 t 5 t ..N/C3
4.6 Technical data
Designation Description Abbr. N C3 Unit
Accuracy class 0.03 0.020 % E
max
Minimum dead load lowest limit of specified measu-
ring range
E
min
0 0 % E
max
Maximum capacity highest limit of specified mea-
suring range
E
max
See Chapter
4.5
Safe load limit maximum load without irrever-
sible damage
E
lim
150 % E
max
Destructive load danger of mechanical destructi-
on
E
d
300 % E
max
Minimum LC verification minimum load cell scale inter-
val, v
min
= E
max
/Y
Y 8333
Minimum preload signal
recurrence
recurrence of the minimum pre-
load signal (DR = ½×E
max
/Z)
Z 3000
Rated output relative output at maximum ca-
pacity
C
n
2 mV/V
Tolerance on rated out-
put
permissible deviation from ra-
ted output C
n
d
c
<0.25 % C
n
Zero output signal load cell output signal under un-
loaded condition
S
min
0 ±2.0 % C
n
Repeatability max. change in load cell output
for repeated loading
ɛ
R
<0.012 <0.012 % C
n
Creep max. change of output signal at
E
max
during 30 minutes
d
cr
<0.030 <0.017 % C
n
Non-linearity
1)
deviation from best straight line
through zero
d
Lin
<0.030 <0.017 % C
n
S-Type Load Cell LC Tigo PR 76 4 Specifications