10. Various functions by SQ function data
10-1. Weighing mode
The measurement mode can be selected by the setting of SQ function SQF-01.
The measurement mode can be selected from the [Simple comparison mode], [Sequential mode],
[4-steps comparator mode 1] or [4-steps comparator mode 2].
Default is set to [Sequential mode].
10-2. Control mode
The control mode can be selected by the setting of SQ function SQF-02.
The control mode can be selected from the [Batch mode], [Discharge mode] or [External input
changeover].
When the [External control changeover] is selected, [Batch mode] is selected when the
[BATCH/DISCHARGE changeover] signal of setting method of the external control input is ON, and
[Discharge mode] is selected when the [BATCH/DISCHARGE changeover] signal of setting method
of the external control input is OFF.
Default is set to [Batch mode]
10-3. Comparison signal operation
The operation condition of the comparison signal can be selected by SQ function SQF-05.
The operation condition of the comparison signal can be selected from [Anytime], [In stable
condition], [Synchronizing with batch finish] or [HOLD at synchronizing with batch finish].
Default is set to [Synchronizing with batch finish].
10-4. Near zero (ZERO BAND) comparison operation
The operation condition of near zero signal can be selected by SQ function SQF-06.
The operation condition of near zero signals can be selected from [GROSS], [NET], [Comparison
signal OFF] or [| GROSS | and | NET |].
Default is set to [GROSS].
10-5. Full comparative operation
The operation condition of [FULL] signal can be selected by SQ function SQF-07.
The operation condition of [FULL] signal can be selected from [GROSS], [NET], [Comparison signal
OFF] or [| GROSS | and | NET |].
Default is set to [GROSS].
10-6. Batch Monitoring Timer
Set the timer that monitors the time from the input of Start signal to the output of Finish signal by
SQ function SQF-10
When the measuring time is longer than the set value, the error is output.
The contents of error are as follows: When the [CONTROL MODE] is selecting [BATCH MODE],
[SQERR 4] can be seen and the [EXTERNAL] is selected, [SQERR 5] can be seen.
Setting range: 0000-9999, Unit: 1s, 0000: OFF
Default is set to [0000]