98
10-15. Setting for the operation of compensation flow
Set for the operation of compensation flow.
10-15-1. Maximum time of compensation flow
Set the maximum time of compensation flow by SQ function SQF-35.
Setting range: 000-255, Unit: 1time, 000:OFF
Default is set to [000].
10-15-2. Judge condition after compensation flow
Judge condition after compensation flow can be selected by SQ function SQF-36.
Judge condition after compensation flow can be selected from [Timer is up], [Stable and timer is up] or
[Stable or timer is up].
Default is set to [Timer is up].
10-15-3. Compensation flow time
The compensation flow time is set by SQ function SQF-37.
The compensation flow time can be set for each brand number by setting SQF-24: Flow/Discharge
brand setting target.
Setting range: 0000 to 9999, Unit: 0.01s, 0000: Compensation flow OFF
Default is set to [0000].
10-15-4. Waiting time for judge after compensation flow
The waiting time for judge after compensation flow is set by SQ function SQF-38.
The waiting time for judge after compensation flow can be set for each brand number by setting
SQF-24: Flow/Discharge brand setting target.
Setting range: 0000 to 9999, Unit: 0.01s, 0000: Compensation flow OFF
Default is set to [0000].
10-16. Setting for the operation of measuring finish output
Set for the operation of measuring finish output
10-16-1. Output time for the Batch finish signal
Set the output time for the Batch finish signal by SQ function SQF-40.
Setting range: 0000-9999, Unit: 0.01s, 0000: OFF
Default is set to [0000].