10-16-2. Batch finish output OFF
The condition to make OFF the output of completion of batch can be selected by SQ function SQF-41.
The condition to make OFF the output of completion of batch can be selected from [No condition], [OL
or unstable] and [Near zero].
When the [No condition] is selected, the FINISH signal will be OFF whether the set time with the
[BATCH FINISH OUTPUT IN] has passed, or next start signal is inputted.
When the [OL or unstable] is selected, the FINISH signal will be OFF whether the display becomes OL
or – OL, or output of Stable signal will become OFF.
When the [Near zero] is selected, the FINISH signal will be OFF at the time of Near zero signals is
outputted.
Default is set to [NO CONDITION].
10-17. Setting for discharge operation after measuring finish
Set for discharge operation after measuring finish.
10-17-1. Discharge start at the time of measuring finish
Set the operation of discharge start at the time of measuring finish by SQ function SQF-45.
Default is set to [Invalid].
10-17-2. Discharge monitor timer
Set the timer that monitors the time from the Discharge start to Discharge finish by SQ function
SQF-46.
Setting range: 0000-9999, Unit: 1s, 0000: OFF
Default is set to [0000].
10-17-3. Discharge gate close delay time
When measurement value fills near zero condition during discharge is executing then set the time
up to gate closing by SQ function SQF-47.
Set the timer that monitors the time from the Discharge start to Discharge finish by SQ function
SQF-46.
Setting range: 0000-9999, Unit: 0.01s, 0000: OFF
Default is set to [0000].
10-17-4. Discharge finish output time
Set the output time for the discharge finish signal by SQ function SQF-48.
Setting range: 0000-9999, Unit: 0.01s, 0000: OFF
Default is set to [0000].
10-17-5. Tare cancellation at the time of discharge finish
Set the operation of tare cancellation at the time of discharge finish by SQ function SQF-49.
Default is set to [Invalid].