189
The external control I/O
is set.
External control input
is wrong.
Connecting wires are
corrected?
Connect surely according to 2-3.
External control I/O check
Yes
No.
No.
Set the external control I/O.
Yes
No.
Set the pulse input value 50ms or more.
Inform Minebea about the contents of
failure and situation in details.
Yes
No.
Yes
Yes
Inform Minebea about the contents of
failure and situation in details.
Set the pulse input value 50ms or more.
No.
50 ms or more
of input signal is set for the
pulse input.
50 ms or more
of input signal for the item of
Level input.
Inform us about the contents of
failure and situation at site in details
Inform us about the contents of
failure and situation at site in details