192
BCD output check
Correct connecting?
No.
Yes
No.
Yes
No.
Yes
No.
Yes
The BCD
output does not change.
The intended
value is not outputted.
Except for
the BCD are out
utted.
Connect wires surely with an external
equipment according to the 16-2-7.
Make the HOLD input opened.
Select the target of BCD output by
referring to the “16-2-3.
Match the setting concerning the BCD
output to the specification of an external
connected equipment according to 16-2-4.
Inform Minebea about the contents of
failure and situation in details.
Inform us about the contents of
failure and situation at site in details