19-2. Error sub display
Error code Contents Measures
In case of execute ZERO out of
range of zero
Check the range of zero is
appropriate.
In case of execute TARE out of
range of tare.
Execute the TARE within the range
of tare.
When the accumulated value and
the accumulated count exceeding
the display range results.
As the accumulating condition
display blinks, please execute to
clear the accumulated and count.
Error display at sub display has been blinking, as long as sequence error is not release.
Press the key to release sequence error.
20. Display of sequence error
Error code Error
No.
Contents Measures
1 Stop compulsorily by inputting
the temporary stop while
weighing
Execute the restart after checking
the source of the temporary stop.
3 When the load value is under
even if the supplementary flow
is executed.
Execute the restart after changing
the supplementary flow time and
numbers.
4 When there is contradiction in
the amount of the comparison
value.
Solve the contradiction of amount
after checking the setting value.
5 When the batching time
exceeds the limited time.
Check the blocking at the batching
gate.
6 When the discharging time
exceed the limited time.
Check the blocking at the
discharging gate.
7 When the gross value is smaller
than the final
value in discharge control.
Add the BRAND.
8 When the net value > (final
value – free fall) in the start
Remove the contents in the
container.
9 When the measurement begins
in the condition that START
ABOVE Near zero is [VALID],
the container is not put on the
measuring section.
Check that whether the container
is put.
Sequence error display has been blinking at sub display, as long as sequence error is not release.
Press the key to release sequence error.