38
6-6. Net weight sign inversion
The net weight sign inversion function is used to invert the sign of the net weight display and net
weight data output externally, for example, in simple comparison discharging mode.
The net weight sign inversion function is a function for making the net weight display or net weight
data output externally positive by inverting the sign of the net weight using C function CF-08.
Net weight sign inversion can be selected from [Net weight sign not inverted] or [Net weight sign
inverted].
Default is set to [Net weight sign not inverted].
6-7. Zero set
The zero set function memorizes a present measuring value as zero point and adjusts the display to
zero by pressing the key when the indicated value of the load display section is the gross value
and at the same time Zero set effective range.
6-7-1. Operating condition of zero set
The operating condition of zero set is selected by C function CF-10.
The effective range is common for both of zero set and zero tracking.
[Operation at steady] or [Unconditional operation] can be selected.
Default is set to [Operation at steady].
6-7-2. Zero set effective range
The effective range of zero set is selected by C function CF-11
Setting range: 00 – 30 Unit : ±1 % of the weighing capacity, 00: OFF
Default is set to [02].
・ When the zero compensation for the effective range is executed by the zero set and the zero
tracking, the zero set is not accepted.
・ When the tare weight cancellation or the preset tare weight cancellation is executed, zero set shall
be executed after clearing the tare weight cancellation and the preset tare weight cancellation.