40
6-9. Power on zero
The power on zero function changes the display to zero if the changes of display is within ±10 % of
the maximum capacity when the power is turn on, or in the stable condition in the operation of
display turning on.
6-9-1. The operation of power on zero
The operation of power on zero is selected by C function CF-15.
The power-on-zero operation is selected from [Effective] and [Invalid].
Default is set to [Invalid].
・ The zero compensation is not accepted in the following condition when the setting of C functions
CF-15 is [Effective power on zero] and when power on zero is starting.
(1) When the display is not a stable condition, it becomes as all lighting.
(2) When the measurement value exceeds ±10 % of the weighing capacity, [ ‐‐‐‐‐‐ ] is displayed.
At this time, by pressing the key, the display becomes a load value compulsorily.
・ When the power on zero is executed, each data of tare weight cancellation, preset tare weight
cancellation, zero set and zero tracking shall be all cleared.
6-10. Tare weight cancellation
By pressing the key, the indication of [TARE] and [NET] in the condition display section is
turned on and the display changes to zero.
To execute the tare weight cancellation clear, press the key when the gross value is zero. The
tare weight cancellation is cleared and [GROSS] of the condition display section lights at the same
time as becoming a gross value display.
6-10-1. Operating condition of tare weight cancellation
The operating condition of tare weight cancellation is selected by C function CF-16.
[Operation when it is steady and 0 < GROSS ≦ WEIGHING CAPACITY],
[Operate unconditionally and 0 < GROSS ≦ WEIGHING CAPACITY],
[Operation when it is steady and 0 < GROSS ≦ | WEIGHING CAPACITY |] and
[Operate unconditionally and 0 < GROSS ≦ | WEIGHING CAPACITY |] can be selected.
Default is set to [Operation when it is steady and 0 < GROSS ≦ WEIGHING CAPACITY]
・ The tare weight cancellation is not accepted when the gross value is larger than the first range.
・ When the tare weight cancellation is executed with a preset tare weight cancellation executed, a
preset tare weight cancellation is canceled, and the tare weight cancellation is executed. (A preset
tare weight cancellation cannot be used together with the tare weight cancellation)
・ When the zero set is executed in the condition of gross value display, the tare weight cancellation
will be cancelled.