70
9-2-2. Measurement comparison value 2 setting mode
The measurement comparison value used in four steps of comparator modes 1 and 2.
(1) Setting method of measurement comparison value 2 setting mode
Change from the standard measurement mode to the
display by the key.
is displayed after pressing the key twice,
and enter into the measurement comparison value 2 setting mode.
By pressing the key, is displayed.
Change this number which you set.
At this time, the corresponding display blinks linked to the display
for S1 to S4, SET7, SET8, and B0 to B7.
: The value of changed digit is changed.
: Changed digit is selected.
: Display returns to .
: The display value is memorized, and it proceeds to the
next step.
Please press the key, and set the setting value of the
selected comparison value.
Please press the key. The setting content is registered and
the display panel displays the following registered measurement
comparison value number.
Please continue and select the selected measurement
comparison value number when you change setting of other
measurement comparison value number.
If the key is pressed after registration is completed, it
becomes display. Please set to the
measurement mode by pressing the key.
Please do not change in the set value under the measurement because the A/D sampling rate
decreases temporarily while rewriting the set value. Please set the storing place to the internal ROM
when you change the set value during the measurement.
Measurement mode
Completing
registration.
Measurement mode