71
(2) Function of the data of measurement comparison value 2
Item Set data No. Set value Contents
S1 S1
-999999999999
000000
Unit : 1D
S2 S2
-999999999999
000000
Unit : 1D
S3 S3
-999999999999
000000
Unit : 1D
S4 S4
-999999999999
000000
Unit : 1D
Near zero SET7
000000999999
000000
Unit : 1D
FULL SET8
000000999999
000000
Unit : 1D
Brand number
Set the brand number.
The brand number can be changed by pressing the key or the key when
is displayed.
Setting range: 0 to 7 Unit: 1D
Default is set to [0].
S1 to S4
Set the value of comparator
Setting range-999999-999999 Unit : 1D
Default is set to [000000].
Operation for value is referring to 9-9. and 9-10.
Near zero (ZERO BAND)
Set the value that can output the signal of near zero.
Setting range000000-999999 Unit : 1D
Default is set to [000000].
When the measurement result is under the value of (fixed valueover) , the under output will be on.
FULL
Set the value of FULL.
Setting range000000-999999 Unit : 1D
Default is set to [000000].
When the Gross weight or the Net weight exceeds the Full value, the FULL output will be ON.
Please set the stored place to the internal RAM when the set value is continuously changed, and the
rewritable time of EEPROM might exceed about one million times.