82
9-4-4. Automatic Free Fall correction
This is a function to guess the next Free Fall from a former actual Free Fall, and to set the value of the
Free Fall automatically when the measurement is operated continuously.
Set data of the next Free Fall uses the moving averaged value calculated from 4 times of actual former
Free Fall values.
When using the function, set the Automatic Free Fall correction to [VALID] and set the value of
[Automatic Free Fall effective width] by referring to the [9-9. Function of the SQ function data].
When the Batch error ( = |measured value - FINAL value|) is exceeded to the Automatic Free Fall
effective width set, it is treated as invalid data, and the Automatic Free Fall calculation is not operated.
When the value of automatic Free Fall effective width is 0, an Automatic Free Fall calculation will not
be operated.
The Automatic Free Fall correction shall be cleared when the change of set value or at the time of
power supply OFF.
Renewing the output of each condition display is synchronizing with A/D sampling.