Type | Autoclave |
---|---|
Model | B FUTURA |
Application | Sterilization |
Temperature Range | up to 134 °C |
Power Supply | 230 V, 50/60 Hz |
Controls device state and component power in standby mode.
Heats chamber and generator before sterilization cycle begins.
Removes air from the chamber via steam shots and vacuum.
Increases chamber pressure to 1.00 bar using steam injection.
Lowers chamber pressure to -0.80 bar using steam and vacuum.
Increases chamber pressure to 1.00 bar for air removal.
Further reduces chamber pressure to -0.80 bar.
Introduces the main sterilization process at high pressure and temperature.
Discharges steam from the chamber and reaches vacuum.
Eliminates moisture and dries the load using negative pressure.
Restores ambient pressure by admitting air into the chamber.
Restores ambient pressure for cycle end and door opening.
Lists error codes, descriptions, and reset modes for system errors.
Lists fault codes, descriptions, and reset modes for component failures.
Lists hazard codes, descriptions, and reset modes for safety issues.
Lists accessory codes, descriptions, and reset modes for external devices.
Details error indications, causes, and solutions for 'E' codes.
Details fault indications, causes, and solutions for 'A' codes.
Details hazard indications, causes, and solutions for 'H' codes.
Details accessory indications, causes, and solutions for 'S' codes.
Procedure for resetting the system on Futura models.
Procedure for resetting the system on Classic models.
Test individual components and functions for troubleshooting and diagnostics.
Select and run specific test cycles for validation.
Reset maintenance cycle counters and related data.
Restore factory default settings and configurations.
Calibrate the PT1 sensor for accurate temperature readings.
Download device statistics, alarm history, and cycle data.
Configure maintenance counters and service intervals.
Manage user accounts, permissions, and access levels.
Configure network settings for remote access and connectivity.
Enter product codes for model-specific configurations.
Store the unique serial number of the autoclave.
Correct differences between displayed and measured pressure values.
Select sterilization times for 134°C cycles.
Start manual or delayed emptying of tanks.
Activate/deactivate backup and overwrite cycle data warning messages.
Reset maintenance/service warning messages based on cycle count.
Test the safety valve via steam pressurisation. Requires password.
Details of the steam generator, its characteristics, and maintenance.
Information on the manifold, solenoid valves, and their operation.
Components and service operations for the door closing and locking system.
Details of pressure transducer and safety switch, including replacement.
Information on the chamber temperature sensor and its calibration.
Procedure for measuring and adjusting door position for proper sealing.
Step-by-step guide for updating the autoclave firmware via USB.