Why are there spots on my Mocom Tethys H10 instruments?
- Rrobert17Aug 4, 2025
Spots on instruments after using Mocom Laboratory Equipment can occur due to several reasons. It could be that the instruments aren't compatible with the thermal disinfection process or the detergent used. Ensure the material is suitable for both. Another reason might be ignoring the salt level; add salt to the integrated water softening system's reservoir. Also, organic or inorganic residues on the instruments can cause spots; position the load correctly, extend the washing phase for particularly dirty items, and manually remove any stubborn dirt. Lastly, check for ferrous residues in the washing chamber and ensure water hardness and iron ion levels are within limits. Clean the basket and accessories as instructed.