76
OPERATOR'S MANUAL
TROUBLESHOOTING
ALARM CODES
The list of alarm codes, the relevant messages displayed and RESET modes, are indicated in
the following table:
ERRORS (CATEGORY A)
CODE
MESSAGE ON THE
DISPLAY
DESCRIPTION
EXIT
A025
DOOR LOCK PROBLEM
Micro-switch state inconsistent with
door state
3 (in cycle)
2 (stand-by)
A070
RESIN REGENERATION
PROBLEM
Lack of water flow from brine to resins
during regeneration phase
3
A116
ADC ERROR
Electronic system ADC error
3 (in cycle)
2 (stand-by)
A120
REFERENCE HEATING
ELEMENT FAULT
Reference heating element acquisition
chain fault.
3 (in cycle)
2 (stand-by)
A121
REFERENCE HEATING
ELEMENT FAULT
Reference heating element acquisition
chain fault.
3 (in cycle)
2 (stand-by)
A122
REFERENCE HEATING
ELEMENT FAULT
Reference heating element acquisition
chain fault.
3 (in cycle)
2 (stand-by)
A123
REFERENCE HEATING
ELEMENT FAULT
Reference heating element acquisition
chain fault.
3 (in cycle)
2 (stand-by)
A124
REFERENCE HEATING
ELEMENT FAULT
Reference heating element acquisition
chain fault.
3 (in cycle)
2 (stand-by)
A125
REFERENCE HEATING
ELEMENT FAULT
Reference heating element acquisition
chain fault.
3 (in cycle)
2 (stand-by)
A500
WATER FILLING PROBLEM
Water flow rate problem during water
filling
3
A501
WATER LEVEL ABOVE
MAX
Water presence above the foreseen
maximum level.
3
A502
WATER FILLING SLOW
Water filling time over the maximum
limit.
3
A503
WATER FILLING FAST
Water filling time over the minimum
limit.
3
A504
AQUASTOP PROBLEM
"Aquastop" solenoid valve (EV3)
malfunction
3
A551
DRAIN PROBLEM
Chamber emptying failed.
3
A552
TANK FULL PROBLEM
Presence of water in the chamber
when not required or level sensor fault.
1
A601
PT1 OPEN
Temperature sensor PT1 failed.
3 (in cycle)
2 (stand-by)
A602
PT2 OPEN
Temperature sensor PT2 failed.
3 (in cycle)
2 (stand-by)
A603
PT3 OPEN
Temperature sensor PT3 failed
3 (in cycle)
2 (stand-by)
A606
MPX INTERRUPTED
Pressure sensor for detection of water
level in the chamber failed.
3 (in cycle)
2 (stand-by)
A611
PT1 SHORT-CIRCUITED
Temperature sensor PT1 short-
circuited.
3 (in cycle)
2 (stand-by)
A612
PT2 SHORT-CIRCUITED
Temperature sensor PT2 short-
circuited.
3 (in cycle)
2 (stand-by)
A613
PT3 SHORT-CIRCUITED
Temperature sensor PT3 short-
circuited
3 (in cycle)
2 (stand-by)
A616
MPX SHORT-CIRCUITED
Pressure sensor for detection of water
level short-circuited
3 (in cycle)
2 (stand-by)
A700
HEATING SYSTEM
PROBLEM
Lack of water heating in the
recirculation system or recirculation
pump failed.
3
A701
DRYING PROBLEM
The fan does not work properly or the
filter is clogged.
3
A702
MISSING HEPA FILTER
It indicates the lack of HEPA filter.
3
A703
DRYING PROBLEM
Drying heater malfunction
2

Table of Contents