What to do if Mocom SUPREME 17 Laboratory Equipment shows a power outage error?
- EErin SaundersOct 31, 2025
If your Mocom Laboratory Equipment displays a power outage error (E000), here are a few things to check: * If there was a power failure, wait for the power to return, then follow the display instructions to open the door, reset the alarm, and repeat the sterilization cycle. * If the autoclave was accidentally turned off, turn it back on, follow the display instructions to open the door, reset the alarm, and repeat the sterilization cycle. * Ensure the power cable is properly connected. If it was disconnected, reconnect it and follow the display instructions. * If the alarm appears at every start-up, ensure the alarm is properly reset. * Check the 16A fuse with a tester and replace it if blown. Then, turn the autoclave on again and follow the display instructions to open the door...








