The following menus can be used:
- SOLENOID VALVES
- OTHER COMPONENTS
These menus are used to test individual component or to
activate and test several components at the same time to
check the different functions that are supposed to be
performed during a sterilization cycle.
- remove steel plate (1) to separate the door cover from the
crossbar/porthole unit
- be careful with the spring/cylinder units (2)
- close the door
- Access the SERVICE menu
- Select COMPONENT TEST
- open EV1 and EV4
- close the other solenoid valves, if open
- activate VP – vacuum pump
With EV5 closed, there should be no air suction from the
middle of the bacteriological filter.
With EV5 open, there should be air suction from the
bacteriological filter.
- de-activate the components after completing the test