Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
28
FUNCTIONING
EN
4.8. 3rd PRESSURE RISE
Alternating vacuum decreases down to the value of - 0.80 bar with 2 pressure rises up to 1.00 bar ensures that the
residual air in the chamber is about 0.06% of the initial total amount.
Following the phases indicated above, the 3
rd
PRESSURE RISE starts, leading to the actual STERILIZATION
PROCESS, carried out at 134°C/2.15 bar or 121°C/1.10 bar depending on the type of cycle selected.
In this pressure rise phase, the activation conditions of the various components are the same as the previous ones.
3
st
PRESSURE RISE - component activation status
STATUS
EV1
High drain
(N.O.)
EV2
Air inlet
(N.C.)
EV3
Low drain
(N.C.)
EV4
Drain
deflector
(3V)
EV5
Chamber
air inlet
(N.C.)
FAN
WATER
PUMP
- EV6
VACUUM
PUMP
CHAMBER TUBE
BUNDLE
HEATING ELEMENT
GENERATOR
HEATING ELEMENT
Part 1
from -
°0.80
to 0.00
Off
open
On
open
Off
closed
On (1)
Off
closed
ON
ON
OFF
ON
ON
Part 2
from 0.00
to 2.14
Off
open
Off
closed
Off
closed
Off (2)
Off
closed
ON
ON
OFF
ON
ON
134° C B Cycle - 4’00’