4.8. 3rd PRESSURE RISE
Alternating vacuum decreases down to the value of - 0.80 bar with 2 pressure rises up to 1.00 bar ensures that the
residual air in the chamber is about 0.06% of the initial total amount.
Following the phases indicated above, the 3
rd
PRESSURE RISE starts, leading to the actual STERILIZATION
PROCESS, carried out at 134°C/2.15 bar or 121°C/1.10 bar depending on the type of cycle selected.
In this pressure rise phase, the activation conditions of the various components are the same as the previous ones.