Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
30
FUNCTIONING
EN
4.9. STERILIZATION PROCESS
The BALANCING phase lasts approximately 15” and allows stabilising the thermodynamic conditions inside the
chamber before starting the actual PROCESS phase.
Maximum, normal and minimum pressure values have been defined for each cycle and must be observed with during
the BALANCING and PROCESS phases.
Cycle type
Pmax bar
Pnorm bar
Pmin bar
121°C
1.14
1.11
1.08
134°C
2.17
2.14
2.12
During BALANCING and PROCESS phases the control system, if necessary, ensures that:
- the chamber tube bundle heating element is powered to increase the T value in the chamber
- EV1 is powered by draining the steam from the chamber if the pressure value exceeds the Pmax permitted value
- water injection pump / EV6 is powered by delivering steam to the chamber if the pressure value is below the Pmin
permitted value
134° C B Cycle - 4’00’