Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING
6. ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING
6.1. ERROR ALARMS - CODE “E”
(**)
1 = OK (warning) - Tap OK on pop-up window.
2 = OK + door unlock / RESET - Keep OK pressed for 3 seconds on the pop-up window and the autoclave will start
the procedure to open the door in full safety. When the RESET key appears in the left area of the LCD, hold it
pressed for 3 seconds.
CODE AND
MESSAGE ON THE
DISPLAY
E000
POWER OUTAGE
CONTACT TECH.
SERVICE
A power voltage failure
occurred
Once the power is back, follow
the indications on the display to
open the door. Reset the alarm
and repeat the sterilization
cycle.
Autoclave accidentally
turned off with the
main switch
Turn the autoclave on again and
follow the indications on the
display to open the door. Reset
the alarm and repeat the
sterilization cycle.
Autoclave accidentally
turned off due to the
disconnection of the
power cable from the
power socket
Reconnect the power cable
properly and follow the
indications on the display to
open the door. Reset the alarm
and repeat the sterilization
cycle.
Alarm shown at every
start-up
Check that the operator has
properly reset the alarm and
give advices on its correct use.
Problem with fuse
holders
Check the mains 16 A fuse
continuity with tester.
Replace the blown fuse with one
having the same rate. Turn the
autoclave on again and follow
the indications on the display to
open the door.
Check the plug of the
fuse holder.
Make sure that the plug is not
excessively tightened.
Tighten the plug correctly by
slightly loosening it.
Problem with the
power cable - cable
check
If the cable is damaged, replace
it.