Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING
6.2. FAULT ALARMS - CODE “A”
(**)
1 = OK (warning) - Tap OK on pop-up window.
2 = OK + door unlock / RESET - Keep OK pressed for 3 seconds on the pop-up window and the autoclave will start
the procedure to open the door in full safety. When the RESET key appears in the left area of the LCD, hold it
pressed for 3 seconds.
CODE AND
MESSAGE ON
THE DISPLAY
A032
FILLING WATER
LEVEL
SENSOR PROBLEM
CONTACT TECH.
SERVICE
Filling tank:
inconsistency of
water level sensor
reading
Sensor connection
problem - check of
connections
Check connections:
- wirings --> level sensors
- sensor wirings --> PCB
Check correct operation of
sensors by means of the
SERVICE/WATER FILLING
TIMES menu
Level reading problem
- check of sensors
Check correct operation of
sensors by means of the
SERVICE/WATER FILLING
TIMES menu.
Replace the damaged level
sensor.
PCB problem during
sensor reading -
reading check
A033
DISCHARGE
WATER LEVEL
SENSOR PROBLEM
CONTACT TECH.
SERVICE
Discharge tank:
inconsistency of
water level sensor
reading
Connector of water
level sensors in the
discharge tank not
connected
Check connections:
- wirings --> level sensors
- sensor wirings --> PCB
Check correct operation of
sensors by means of the
SERVICE menu
Failure of the water
level sensor(s) in the
discharge tank
Check correct operation of
sensors by means of the
SERVICE menu.
Replace the damaged level
sensor.
PCB problem during
sensor reading -
reading check
A034
FILLING TANK
FILTER
REMOVED DURING
CYCLE
Filling tank filter
removed during
cycle
Water filling tank filter
removed during cycle
Refit the integrated filter in the
filling tank
Failure of filling tank
filter presence sensor
Check correct operation of the
sensor by means of the SERVICE
menu.
Replace the damaged sensor.
PCB problem during
sensor reading -
reading check