Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
EN
ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING
81
CODE AND
MESSAGE ON
THE DISPLAY
DESCRIPTION
Possible causes /
Checks
Solution
ERROR
OUTPUT
(**)
A035
DISCHARGE TANK
FILTER
REMOVED DURING
CYCLE
Discharge tank
filter removed
during cycle
Water discharge tank
filter removed during
cycle
Refit the integrated filter in the
discharge tank
2
Failure of discharge
tank filter presence
sensor
Check correct operation of the
sensor by means of the SERVICE
menu.
Replace the damaged sensor.
PCB problem during
sensor reading -
reading check
Replace the PCB
A040
FAILED WATER
INLET
CHECK
AUTOMATIC
FILLING
Tank filling failed
with automatic
filling
For connection of
external filling pump -
operation check
Check connections:
- pump power cable
-->jack socket on the rear side
of the autoclave
- pump outlet pipe --> automatic
filling connection on the
autoclave rear side
Check:
- pump suction filter
- activation of external pump by
means of the SERVICE/WATER
FILLING TIMES menu.
1
For connection of
demineralizer PURE
100/ PURE 500
Check connections:
- solenoid valve power cable--
>jack socket on the rear side of
the autoclave
- demineralizer outlet pipe -->
automatic filling connection on
the autoclave rear side
Check:
- water mains supply
- Inspectionable filter on the
autoclave/demineralizer
connection
- activation of demineralizer
solenoid valve by means of the
SERVICE/WATER FILLING
TIMES menu
Level sensor problem -
sensor check.
Check:
- minimum level sensor wiring
- correct operation of sensors by
means of the SERVICE/WATER
FILLING TIMES menu
Replace the level sensor