Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
EN
SERVICE MENU
53
5.2.1. COMPONENT TEST FUNCTIONAL CHECKS ON SOLENOID VALVES/OTHER
COMPONENTS
SQ9
The following menus can be used:
- SOLENOID VALVES
- OTHER COMPONENTS
These menus are used to test
individual component or to activate
and test several components at the
same time to check the different
functions that are supposed to be
performed during a sterilization
cycle.
To activate several components
available in the two menus at the
same time, you may toggle
between the two menus using the
command
/
which changes between menus
while retaining the previously
selected commands.