Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
10
GUIDELINES - SAFETY TEST - VALIDATION (IF REQUIRED)
EN
3.2. CONFORMITY REPORT
CONFORMITY REPORT - Ref. Standard:
STANDARD
CLASS
Scheduled maintenance
IEC 61010
CL1
DEALER:
Unscheduled maintenance
DENTAL SURGERY Dr.:
DEVICE:
SERIAL NUMBER:
MEASUREMENT INSTRUMENT - MODEL:
S/N:
VISUAL INSPECTION
Conditions of:
power cables and connectors
earth cables and correct tightening
electrical protections
covers
warning plates
light indicators
OK
Presence of:
liquid leaking out from ducts or on the bottom of the device
user manual
ELECTRICAL SAFETY TEST
(MEASUREMENT POINT AND MAX VALUE)
MEASUREMENT OF EARTH CONDUCTOR RESISTANCE
Ω
OK
Earth conductor ≤ 0.1 Ω
MEASUREMENT OF EARTH-LEAKAGE CURRENT:
A
OK
1) under normal conditions ≤ 0.5 mA
2) under single failure conditions ≤ 3.5 mA
INSULATION RESISTANCE TEST - Test voltage = 500 Vdc
Ω
OK
Measurement between network and earth > 2 MΩ
VISUAL INSPECTION
Door opening and closing controls
Light indications and/or acoustic signals
OK
Notes:
FINAL EVALUATION:
OK
FAULT
DEALER:
TECHNICIAN:
DATE:
SIGNATURE: