Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
32
FUNCTIONING
EN
4.10. DISCHARGE
At the end of the PROCESS cycle, the DRAINAGE phase starts to allow draining all the steam present in the
chamber, with a vacuum decrease to the value of - 0.80 bar.
As for previous VACUUM DECREASE phases, also the DRAINAGE includes a first spontaneous depressurization
that exploits the pressure generated by the steam and that only requires the control of drainage solenoid valves EV1
and EV3.
To complete the DRAINAGE until reaching the value of 0.50 bar, the decrease continues by means of the vacuum
pump and the simultaneous activation of solenoid valve EV4.
In this phase too, activations of solenoid valve EV2 are required to balance the pressure value inside the hydraulic
circuit and facilitate the activation of the vacuum pump.
DRAINAGE - component activation status
STATUS
EV1
High
drain
(N.O.)
EV2
Air inlet
(N.C.)
EV3
Low
drain
(N.C.)
EV4
Drain
deflector
(3V)
EV5
Chamber
air inlet
(N.C.)
FAN
WATER
PUMP EV6
VACUUM
PUMP
CHAMBER TUBE
BUNDLE
HEATING ELEMENT
GENERATOR
HEATING ELEMENT
Direct steam
drainage
Off
open
Off
closed
Off
closed
Off (2)
Off
closed
ON
OFF
ON
ON
ON
Direct condensate
drain
On
closed
Off
closed
On
open
Off (2)
Off
closed
ON
OFF
ON
ON
ON
Decrease from 0.10
to - 0.50 bar
On
closed
Off
closed
On
open
On (1)
Off
closed
ON
OFF
ON
ON
ON
Rise from - 0.50
to - 0.20 bar
On
closed
Off
closed
On
open
On (1)
ON
open
ON
OFF
ON
ON
ON
Decrease from -0.20
to - 0.80 bar
On
closed
Off
closed
On
open
On (1)
OFF
closed
ON
OFF
ON
ON
ON
134° C B Cycle - 4’00’