ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING
CODE AND
MESSAGE ON THE
DISPLAY
E001
OVERVOLTAGE
CONTACT TECH.
SERVICE
Supply voltage value
beyond limits - check
supply voltage
Disconnect the power cable and
check the mains voltage. The
value must be within the rated
voltage limits +/- 10%.
If it exceeds the limits, do not
connect the autoclave and
check the power mains.
Excessive voltage
value at transformer
output - check output
voltage
Disconnect the wiring at
transformer output and check
voltage. The value must not
exceed 34V.
Replace the transformer.
E002
INSUFFICIENT H
2
O
QUALITY
H
2
O conductivity
above 20 uS/cm
(manual filling
without integrated
filters)
Use of poor quality
demineralised water -
check label data
Empty the filling tank and fill it
with demineralised water falling
within the limit of 15 μS.
Conductivity sensor
problem
Check that the water in the tank
has an appropriate conductivity
using an external sensor. Empty
the filling tank, replace the
conductivity sensor and fill it
with demineralised water falling
within the limit of 15 μS.
E003
QUALITY H
2
O BAD
CHANGE WATER
H
2
O conductivity
above 60 uS/cm
(manual filling
without integrated
filters)
NOTE:
IN THESE
CONDITIONS, THE
DEVICE ALLOWS
STARTING UP TO 5
CONSECUTIVE
CYCLES, THEN IT
LOCKS OUT UNTIL
WATER WITH
PROPER QUALITY
IS DETECTED (<15
μS/cm).
THIS PRECAUTION
IS NECESSARY TO
PREVENT ANY
DAMAGE TO THE
DEVICE.
Use of water with
inadequate quality.
Lock counter activated
- it is possible to run
up to 5 cycles.
Empty the filling tank and fill it
with demineralised water falling
within the limit of 15 μS.
Activate the sterilization cycle to
reset the cycle counter for
autoclave lock.
1/2
(versions
Futura /
Supreme)
Conductivity sensor
problem
Check that the water in the tank
has an appropriate conductivity
using an external sensor. Empty
the filling tank, replace the
conductivity sensor and fill it
with demineralised water falling
within the limit of 15 μS. Activate
the sterilization cycle to reset
the cycle counter for autoclave
lock.