Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
70
ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING
EN
CODE AND
MESSAGE ON THE
DISPLAY
DESCRIPTION
Possible causes /
Checks
Solution
ERROR
OUTPUT (**)
E004
LINE FREQ. ERROR
CONTACT TECH.
SERVICE
Mains frequency
reading error
Electrical mains failure
RESET following the
instructions.
Connect the device to another
power socket.
If the problem persists, have the
mains system checked by an
electrical installer (check also
the uninterruptible power
system, if present).
Main board fault
RESET following the
instructions.
Replace the board.
E007
FAN PROBLEM
CONTACT TECH.
SERVICE
One of the two fans
does not rotate
Breakage of one or
more rear fans
RESET following the
instructions.
Check the operation of rear
fans.
E008
FILTERS NEARLY
EXHAUSTED
H
2
O conductivity
above 20 uS/cm
(automatic filling with
centralised system
or Pure 100/500, or,
with integrated filters
and water filling from
filling tank)
For connection to
demineralizer PURE
100 - device check
Empty the filling tank, replace
the 2 PURE 100 resin
cartridges.
Top up the tank.
1
For connection to
demineralizer PURE
500 device check
Empty the filling tank, replace
the PURE 500 filtering element
kit.
Top up the tank.
For connection to
demineralization
centralised system -
check the quality of the
water supplied by the
system.
Empty the filling tank, select
manual supply mode and fill the
tank with demineralised water
falling within the limit of 15 μS.
After checking and restoring the
correct conductivity value of the
water supplied by the
centralised system, select the
automatic supply and top up the
tank.
Conductivity sensor
problem
Check that the water in the tank
has an appropriate conductivity
using an external sensor. Empty
the filling tank, replace the
conductivity sensor and fill it
with demineralised water falling
within the limit of 15 μS.