Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
EN
ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING
71
CODE AND
MESSAGE ON THE
DISPLAY
DESCRIPTION
Possible causes /
Checks
Solution
ERROR
OUTPUT (**)
E009
QUALITY H
2
O BAD
CHANGE FILTERS
CHANGE WATER
H
2
O conductivity
above 60 uS/cm
(automatic filling with
centralised system
or Pure 100/500, or,
with integrated filters
and water filling from
filling tank)
NOTE:
IN THESE
CONDITIONS, THE
DEVICE ALLOWS
STARTING UP TO 5
CONSECUTIVE
CYCLES, THEN IT
LOCKS OUT UNTIL
WATER WITH
PROPER QUALITY
IS DETECTED (<15
μS/cm). THIS
PRECAUTION IS
NECESSARY TO
PREVENT ANY
DAMAGE TO THE
DEVICE.
For connection to
demineralizer PURE
100 device check.
Empty the filling tank, replace
the 2 PURE 100 resin
cartridges.
Top up the tank, activate the
sterilization cycle to reset the
cycle counter for autoclave lock.
1/2
(versions
Futura /
Supreme)
For connection to
demineralizer PURE
500 - device check
Empty the filling tank, replace
the PURE 500 filtering element
kit.
Top up the tank, activate the
sterilization cycle to reset the
cycle counter for autoclave lock.
For connection to
demineralization
centralised system -
check the quality of the
water supplied by the
system.
Empty the filling tank, select
manual supply mode and fill the
tank with demineralised water
falling within the limit of 15 μS.
Check and restore the correct
conductivity value of the water
supplied by the centralised
system, select automatic supply,
top up the tank, activate the
sterilization cycle to reset the
cycle counter for autoclave lock.
Conductivity sensor
problem.
Check that the water in the tank
has an appropriate conductivity
using an external sensor. Empty
the filling tank, replace the
conductivity sensor and fill it
with demineralised water falling
within the limit of 15 μS. Activate
the sterilization cycle to reset
the cycle counter for autoclave
lock.
E010
DOOR OPEN
CLOSE DOOR "
Door open
The door is not
properly closed.
Properly train the operator on
the correct ways to close and
open the door.
1
Door closing activation
failure - check ajar
door microswitch.
Check the correct operation of
the microswitch by means of the
SERVICE/COMPONENT
TEST/DOOR menu.
Replace the microswitch - see
the technical sheet DOOR
CLOSING UNIT.