Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
72
ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING
EN
CODE AND
MESSAGE ON THE
DISPLAY
DESCRIPTION
Possible causes /
Checks
Solution
ERROR
OUTPUT (**)
E011
QUALITY H
2
O BAD
CHANGE FILTERS
CHANGE WATER
H
2
O conductivity
above 60 uS/cm
(only with integrated
filters and filling from
discharge tank)
Very poor quality water
in the discharge tank.
RESET following the
instructions.
IMMEDIATELY empty both
tanks, replace filtering elements
and fill the filling tank.
Activate the sterilization cycle to
reset the cycle counter for
autoclave lock.
2
Conductivity sensor
problem.
Check that the water in the tank
has an appropriate conductivity
using an external sensor.
Empty the filling tank, replace
the conductivity sensor and fill it
with demineralised water falling
within the limit of 15 μS.
Activate the sterilization cycle to
reset the cycle counter for
autoclave lock.
E012
CHANGING FILTERS
The maximum
number of cycles
that can be run with
filters has been
reached or the water
conductivity level is
too high (only with
integrated filters)
The number of cycles
after which integrated
filters must be
replaced has been
reached.
RESET following the
instructions. IMMEDIATELY
empty both tanks, replace
filtering elements and fill the
filling tank.