Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
EN
ALARMS / ERRORS / WARNINGS / TROUBLESHOOTING
73
CODE AND
MESSAGE ON THE
DISPLAY
DESCRIPTION
Possible causes /
Checks
Solution
ERROR
OUTPUT (**)
E013
QUALITY H
2
O BAD
CHANGE FILTERS
CHANGE WATER
The maximum limit
of 5 cycles run with
H2O conductivity
above 60 uS/cm has
been reached.
For connection to
demineralizer PURE
100 device check
Empty the filling tank, replace
the 2 PURE 100 resin cartridges
and the integrated
demineralization filter.
Top up the tank, activate the
sterilization cycle to reset the
cycle counter for autoclave lock.
2
For connection to
demineralizer PURE
500 - device check
Empty the filling tank, replace
the PURE 500 filtering element
kit and the integrated
demineralization filter.
Top up the tank, activate the
sterilization cycle to reset the
cycle counter for autoclave lock.
For connection to
demineralization
centralised system -
check the quality of the
water supplied by the
system
Empty the filling tank, replace
the integrated demineralization
filter.
Select manual supply mode and
fill the tank with demineralised
water falling within the limit of 15
μS.
After checking and restoring the
correct conductivity value of the
water supplied by the
centralised system,
- select the automatic supply
- top up the tank.
- activate the sterilization cycle
to reset the cycle counter for
autoclave lock.
For models with
integrated filtering
elements (only
demineralizer,
demineralizer +
recirculation)
- empty both tanks
- replace filtering elements
- fill the filling tank
- activate the sterilization cycle
to reset the lock counter
Conductivity sensor
problem
Check that the water in the tank
has an appropriate conductivity
using an external sensor.
Empty the filling tank, replace
the conductivity sensor and fill it
with demineralised water falling
within the limit of 15 μS.
Activate the sterilization cycle to
reset the cycle counter for
autoclave lock.