EasyManua.ls Logo

Mocom Tethys H10 - ERRORS (CATEGORY S)

Mocom Tethys H10
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
80
OPERATOR'S MANUAL
ERRORS (CATEGORY S)
CODE
MESSAGE ON THE
DISPLAY
DESCRIPTION
EXIT
S001
FLASH NOT ACCESSIBLE
Flash memory not accessible
2
S002
FLASH FULL
Flash memory full.
2
S003
SD CARD NOT
ACCESSIBLE
Micro SD card not accessible for data
storage in start-up phase.
2
S004
SD CARD FULL
Micro SD Card full
2
S005
USB STICK ERROR
USB stick not accessible (interruption
of communication after download
started successfully).
2
S006
USB STICK NOT
ACCESSIBLE
USB Stick not accessible (before
download start)
2
S007
USB STICK FULL
USB stick full
2
S008
SD CARD NOT
ACCESSIBLE
Micro SD not accessible during data
storage phase
2
S018
IMPOSSIBLE TO OPEN
CRC FILE
CRC file inside SD card cannot be
opened
2
S020
RUN BACKUP
Cycle report to be downloaded
recommended limit has been reached.
2
S021
OVERWRITE DATA
Cycle report to be downloaded limit in
case of SD Card malfunction has been
reached.
2
S030
SYSTEM ERROR
Malfunction in software running.
2 (out of
cycle)
3 (in cycle)
S031
SYSTEM ERROR
Malfunction in software running.
2 (out of
cycle)
3 (in cycle)
S032
SYSTEM ERROR
Malfunction in software running.
2 (out of
cycle)
3 (in cycle)
S033
SYSTEM ERROR
Malfunction in software running
2 (out of
cycle)
3 (in cycle)
S100
INCONSISTENT STATUS
Software problem in management of
the correct sequence of events
3
1 = OK (warning) + chamber emptying
2 = OK (warning)
3 = cycle interruption + ok + door unlock + reset 3

Table of Contents