19
10.0 SPECIFICATIONS
Note: Product specificaons are subject to change without noce.
10.1 Measurement
Pressure Ranges: Various pressure ranges absolute and
gauge. Check ordering informaon
Accuracy: ±0.25% Full Scale typical (incl. linearity,
repeatability and hysteresis)
Resoluon: 0.01%. Display 4½ Digits. Max 2 decimal
places.
Compensated Range -10°C to +70°C typical
Temperature Coeff. 0.02% Full Scale/°C. (> 580 PSI or 4MPa)
0.03% Full Scale/°C. (< 580 PSI or 4MPa)
Overpressure Rang: 1.5 mes max
Port connecon: ¼” or ½” NPT Male
Sensor Housing & Diaphragm: Stainless Steel 316L Fill: Silicon Oil
Temperature Range: -20 to +85°C / -4 to 185°F
Accuracy: ± 5°C
Resoluon: 0.1°C / 0.2°F
10.2 General
Record: Sample Rates: User configured 1 every 2 seconds
up to 1 every 24 hours
No. of Samples: 130,000 (depends on setup)
w ww . . co m