POLI User’s Guide
13
Calibration is required if:
The sensor module has been replaced with one whose calibration is overdue.
The defined period of time between calibrations has been exceeded.
Users have changed the calibration gas type without recalibrating the instrument.
The sensor has failed in a previous calibration.
4.6 Bump Status
If one or more sensors requires a bump test, then the screen displays the word ‘BUMP’ at
the place of the sensor name, and alternates between the sensor name and the word
Bump with a highlighted background:
A bump test is required if the defined period of time between bump tests has been
exceeded. This interval is set by an administrator using mPower Suite software.