MRC Ltd
32 http://www.mrclab.com/
Fuse of socket and fuse of neutral wire is located on top of the equipment (refer
to picture 13). For replacing the fuse, turn off the power and disconnect the plug.
Use a Phillips screwdriver and rotate it anticlockwise to unscrew the fuse holder.
Replace the fuse inside the fuse holder and then, use a Phillips screwdriver and
rotate it clockwise to screw back the fuse holder. Live wire fuse is also located
on top of the equipment, inside the female connector. For replacing the live wire
fuse, turn off the power and disconnect the plug. Use a Slotted screwdriver to
lever up the fuse holder to open it. Replace the fuse inside the fuse holder and
then, press the fuse holder back.
The parameter of the fuse tubes in the round fuse holders are required to
conform to labels, that are F5A φ5×20 mm and F10A φ5×20 mm.
Picture 16
b) Replace fluorescent light
For replacing the lamp tube, press UP button to rise up the front window to the
highest position, then turn off the power and disconnect the plug. Remove the
power cord of the lamp tube refer to picture 17. Then remove the lamp tube by
pulling the small slice on the tube base as shown in the Picture 17. Install a new
lamp tube by pushing it in and connecting with the power cord.