MRC Ltd
1. Caster
2. Base cabinet
3. Water sink
4. Water tap
5. Water tap control knob
6. Gas tap control knob
7. Access panel
8. Control panel
9. Gas tap
10. Power socket
11. Fuse holder
12. Exhaust outlet
13. Front window
14. Phenolic compact laminate
15. Work table
16. Socket
17. Gas inlet
18. Water inlet
19. Drain hose
2.2.2. Structure introduction
a) Driving system of front window
Driving system consists of tubular motor, front window and hauling mechanism
(hauling sash)
b) UV lamp
The entire work zone could be sterilized effectively by the UV lamp located at
the top of work zone. Emission of 253.7 nanometers could ensure the most
efficient decontamination.
c) Fluorescent light
The Fume Hood is equipped with LED lamp tube, which ensures the standard
requirement of average illumination (400 lux) is met. The measured value at any
point inside the working zone should be greater than 350 lux.