Front window down, press and hold to continuously lower the window
Socket power, press to activate the sockets in the working zone
Adjust fan speed, press to adjust the speed of the blower (fan) from F1 to F9
2.3.2. Control of the front window
The height of the front window could be adjusted by pressing the and
buttons. The window will be moved continuously when pressing the button and it will
stop moving immediately when the button is released
Please make sure the height of the front window is within the safety height range (520
mm).
2.3.3. Control of the fan speed
The blower (fan) could be turned on by pressing button (only when the UV lamp
is turned off). The speed of the fan could be adjusted by pressing button. 9 speed
levels could be selected. The relevant speed level would be displayed on the LED
screen.
When turn off the blower or power off the Fume Hood, the level of fan speed is
memorized by the equipment and would be resumed when turn on the blower again.
2.3.4. UV sterilization
UV light could be turned on by pressing the UV button. Please make sure the window is
fully closed before starting UV sterilization. Interlock function was adopted between
UV light and the blower/fluorescent light. UV light could be automatically turned off
when either the blower or fluorescent light is turned on. During sterilization, people
should leave the room for safety of eyes and skin.
2.4. Instructions of Operation
2.4.1. Normal Operation Notice
a) Make
sure input voltage is correct and stable. The rated load of main power
socket should be higher than cabinet consumption. Plug must be well grounded.