b) The equipment should be powered off and unplugged before doing any
replacement of parts, such as UV lamp and fluorescent lamp.
c) The front window is made of explosion-proof toughened glass. In order to keep
the front window clean and clear, please wipe it by wet soft cloth and keep it
away from hydrofluoric acid
d) The air deflector and other internal accessories should be cleaned according to
the use of the Fume Hood
e) The air duct and the blower of the Fume Hood should be cleaned and maintained
regularly in a proper way
f) Fume Hood should be placed in a position where there should be no other
equipment or machine within 150mm of the front window
g) Do NOT place any soft or tiny materials (such as soft tissue) on the work table
during the operation to prevent breakdown of the blower causing by sucking
those materials
h) The packed Fume Hood should be stored in a warehouse with relative humidity
no more than 75% and temperature lower than 40. The warehouse should have
good ventilation performance without acid, alkali or other corrosive gases
i) The maximum storage period is one year. A performance inspection should be
done if the storage period exceeds one year
NOTE: MRC WILL NOT BE LIABLE FOR ANY RISK OR
DAMAGE ARISING FROM YOUR FAILURE TO APPROPRIATELY
OPERATION THE FUME HOOD!
2.4.2. Operation Process
a) Connect to a suitable power supply
b) Power on the Fume Hood by pressing the power switch under the working zone,
the LED screen would be lighted as “
c) Press the POWER button on the control panel to enable all functions
(fluorescent lamp, UV lamp, blower, socket, front window). The LED screen